Welcome to the
TAMU MINOS work area
Measurements of Z Positions of Planes
Here are the links to the measurements of the Z-positions of the face of
the steel plates near the axial bolts at the far detector. If you have
any questions about these measurements please let me know.
Plane Checkout Procedures
Here is the latest version of the plane checkout QC sheet. At this time
there is a single set of sheets outlining the procedures done both horizontally
and vertically. This is the form as it was modified following the Installation
Workshop at Tower on 10/13-14/2001.
LeCroy 1440 HV system parameters for MINOS detector operation
Below you will find figures referencing several sets of measurements that
we made on a LeCroy 1440 here at TAMU. These measurements were made using
a phototube base drawing 0.5 mA of current on the output being tested.
This is similar to the load that we will be using on the M16 base planned
for the far detector.
This figure show the effect of the hardware HV limit setting on the voltage
being delivered to the base. Measurements were taken with three values
of the HV limit setting; 0.3, 0.35 and 0.4. At each setting we then ramped
the requested voltages up in steps and "measured" the voltage detected
on the output with a DVM. We saw that at the 0.3 setting we could get to
an output voltage of 1000 volts but this required asking the supply to
put out a voltage of 1200 volts or so to reach this level. If we have tubes
that will require operating at the HV limit of 1000 volts we will either
have to operate the system with a slightly higher HV limit like 0.35 or
be prepared to allow for demand voltages higher than 1000 volts to be accepted
and processed by the DCS system.
This figure shows the measured output voltage as a function of the current
limit setting of the 1440 system. To make these measurements, we fixed
the requested voltage at 1000V\ and then changed the current limit from
0 - 256 and at each current limit setting we measured the voltage being
delivered to the test base. In this figure we display the voltages measured
by DVM as well as the voltages readback by the 1440 system itself.
In this figure, we carry out the same measurements as described in the
paragraph above. However this time the figure displays two different fixed
values of the requested output voltage and the response to the system while
varying the current limit setting.
From these two sets of measurements it looks like the 1440 systems should
be run with the hardware HV limit set at between 0.3 and 0.35 AND for an
extra measure of protection we should probably set the current limit to
be in the range of 100-125.
Cross talk measurements on 2 M16s (New and Old)
Below you will find figures referencing the cross-talk measurements that
we have made here at TAMU. In the case of the Old M16, we saw that there
were some curious cross-talk patterns in the tube when certain pixels were
illuminated. An example of this is shown in the ADC spectra for the 16
pixels when pixel 6 is illuminated. All of the surrounding cells clearly
show a single photoelectron peak, and cell 14 has many more hits than cell
4.
We performed similar measurements on a newer MINOS M16(SN # 9H25C2) and
saw a much nicer behavior. When illuminating pixel 6 in this tube we saw
that the cross-talk was independent of the illumination of the primary
pixel and that there were no anomalous pixels. At 800 V we measured the
cross-talk and found that it was hard to see whether the cross-talk signals
were really 1 pe or a continuous charge distribution, see
To investigate this further, we took the same set of data at a voltage
of 1000 V to see if we could see nice pe distributions. The results of
these measurements are shown in
In this plot you can clearly see that the signals in the cross-talk pixels
show strong evidence of a 1 pe peak. Lastly, we summarized these measurements
over the full range of illuminations in
In our cross-talk measurements on these M16s we saw no illumination dependences
of the cros-talk and the newer M16 had a
as shown in the figure. In this plot the vertical bars represent the cross-talk
measured for that particular pixel as a percentage of the light detected
in the primary pixel. Each of these groups represents an illumination at
a different number of primary pes.
Far Detector Installation Issues(all are plain text files)
Results of Split Cookie Construction
Information for Harness Installation on 4PPT
Calibration and Cross Talk measurements
Created January 12, 1998 - last update March 21, 2001
Robert C. Webb webb@heprb.physics.tamu.edu